Anaes Vivas · Measurement & monitoring physics
Pressure and temperature measurement — physics viva
A structured oral covering: pressure as force per unit area with the unit conversions (1 kPa = 7.5 mmHg = 10.2 cmH2O; 1 atm = 101.325 kPa = 760 mmHg); the strain-gauge piezoresistive transducer (diaphragm, four strain gauges, Wheatstone bridge, voltage proportional to pressure); the arterial line components (cannula, non-compliant tubing, pressurised heparinised-saline flush, transducer, monitor); zeroing to atmosphere and levelling to the mid-axillary line with the hydrostatic error of 0.75 to 1 mmHg per cm; the natural frequency and damping (under/over/optimal) and the fast-flush square-wave test with mean preserved in all states; the CVP a/c/v waves and x/y descents with ECG timing and the pathological patterns; the PA catheter pressures at 30/40/50/60 cm and the wedge approximating left atrial pressure; the four temperature principles (thermistor negative coefficient, thermocouple Seebeck voltage, RTD positive coefficient, infrared); and the monitoring sites. The under-damping data probe tests the recognition that a ringing waveform gives a spurious high systolic and that the mean is the trustworthy value.
On this page
Study tools
Target exams
Structure of the viva and expected answers
Opening (2 minutes) — pressure, units and the transducer
The examiner opens with: "Define pressure and tell me how invasive blood pressure is measured." A model opening:
"Pressure is force per unit area; the SI unit is the pascal. Clinically we use mmHg, cmH2O, kPa, bar and atmosphere, with standard interconversions — see a reference table for the exact factors. Most clinical pressures are GAUGE pressures, referenced to atmosphere."[1]
"Invasive blood pressure is measured by a pressure transducer — a device that converts the hydraulic pressure of a fluid column into an electrical signal."[1]
The arterial line and zeroing and levelling (3 minutes)
The examiner probes: "What is in the line, and what two setup steps must you do?" The candidate answers:
"The line runs from the arterial cannula through short, stiff, non-compliant pressure tubing, a continuous-flush device, the transducer, to the monitor. The flush bag is pressurised well above systolic pressure and delivers a slow constant heparinised-saline flush.[3]"
"The two mandatory steps are zeroing (open to atmosphere, defining zero) and levelling (align the transducer with the right atrium). Both are systematic errors — re-level whenever you move the table or the patient.[5]"
Probe: "Why is the flush bag pressurised?" Answer: "So the flow through the restrictor stays slow and constant regardless of the patient's arterial pressure, keeping the column moving and opposing clotting at the tip."
[3]Damping and the fast-flush test (4 minutes)
References6ShowHide
- [1]Gardner RM Direct blood pressure measurement--dynamic response requirements. Anesthesiology, 1981.PMID 7469106
- [2]Kleinman B, Powell S, Kumar P The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology, 1992.PMID 1466471
- [3]Scheer B, Perel A, Pfeiffer UJ Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Critical Care, 2002.PMID 12133178
- [4]Sessler DI Temperature monitoring and perioperative thermoregulation. Anesthesiology, 2008.PMID 18648241
- [5]Pickering TG, Hall JE, Appel LJ, et al Recommendations for blood pressure measurement in humans: an AHA scientific statement from the Council on High Blood Pressure Research Professional and Public Education Subcommittee. J Clin Hypertens (Greenwich), 2005.PMID 15722655
- [6]Harvey S, Harrison DA, Singer M, et al Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet, 2005.PMID 16084255