Anaes Vivas · Measurement & monitoring physics
Data viva — humidification devices and perioperative heat loss
A data viva on humidification and heat — interpreting a diagram that contrasts the passive HME (recycles the patient's own exhaled heat and water on a hygroscopic membrane, single-use, low resistance, about 25 to 32 mg/L) with the active heated water-bath humidifier (warms a reservoir and saturates the gas, near-physiological humidity, for long-term ventilation), and relating them to the four routes of perioperative heat loss dominated by radiation at about 60 per cent.
On this page
Study tools
Target exams
Stimulus
A two-panel diagram with a pie chart beneath it. The left panel shows a passive heat and moisture exchanger: a small housing containing a hygroscopic membrane, interposed between the tracheal tube and the breathing circuit. The right panel shows an active heated water-bath humidifier: a water reservoir warmed by a heater plate, with a heated wire running along the inspiratory limb and a temperature probe sited at the patient's airway. The pie chart below divides perioperative heat loss into four sectors — radiation the largest at about 60 per cent, then convection, then evaporation, then conduction the smallest.