Equipment viva — breathing systems, circle checks and absorbent safety
An equipment viva on breathing systems — the per-system per-mode fresh gas flows (Magill at minute ventilation, Bain at 1.5 to 2 times, T-piece at 2.5 to 3 times spontaneous), the Bain inner-tube tests with the under-20-kg exclusion, the AAGBI two-bag and 20 to 60 cmH2O leak tests, and the strong-base absorbent story with the inspired-CO2 change rule.
On this page
Study tools
Target exams
ANZCAFRCAABAEDAICFCAIFCA_SA
Prompt
A Bain circuit, a Jackson-Rees T-piece and a circle absorber assembly on the viva table. Compare the Mapleson systems by efficiency, demonstrate the circle leak checks, and explain when the absorbent must be changed.
Stimulus
A Bain circuit, a Jackson-Rees T-piece and a circle absorber assembly on the viva table. For each: name the arrangement, state the flow that prevents rebreathing, and show the bedside checks. Finish with when the absorbent gets changed.