Anaes SAQs · Vaporisers
The desflurane vaporiser
Primary physics — desflurane's near-atmospheric saturated vapour pressure (669 mmHg) and low boiling point (22.8 degrees) demand a heated, pressurised, electronically-metered vaporiser.
On this page
Study tools
Target exams
ANZCAFRCAABAEDAIC
Prompt
Explain why desflurane requires a special vaporiser, and describe the design and the operation of the desflurane vaporiser.
Question
Explain why desflurane requires a special vaporiser, and describe the design and the operation of the desflurane vaporiser. (Question stem restated — the worked answers below carry the primary-source citations.)