Anaes SAQs · Measurement & monitoring physics
Pressure transduction, the arterial line, damping, and temperature measurement
A First-examination physics-and-measurement SAQ testing the full invasive-pressure and temperature-measurement framework: pressure as force per unit area with unit conversions; the strain-gauge (piezoresistive) transducer (diaphragm, strain gauges, Wheatstone bridge); the arterial line components (cannula, non-compliant tubing, pressurised heparinised-saline flush, transducer, monitor); zeroing to atmosphere and levelling to the mid-axillary line (right atrium) with the hydrostatic error of about 0.75 to 1 mmHg per cm; the natural frequency and damping (under/over/optimal) and the fast-flush square-wave test with what each state does to systolic, diastolic and mean; the four temperature-sensing principles (thermistor negative coefficient, thermocouple Seebeck voltage, RTD positive coefficient, infrared); and the monitoring sites (oesophageal/nasopharyngeal/pulmonary artery/bladder core vs skin/peripheral).
On this page
Study tools
Target exams
Model answer plan (20 marks, 20 minutes)
A high-scoring answer is structured in five blocks that mirror the stem: (1) principles and units; (2) the transducer and the arterial line; (3) zeroing, levelling and the dynamic response; (4) temperature principles; (5) monitoring sites. Allocate roughly 4 marks to each block. [1]
1. Principles and units (4 marks)
- Pressure is force per unit area (P = F/A); SI unit the pascal (Pa = N/m2).
- Conversions (state at least three): 1 kPa = 7.5 mmHg = 10.2 cmH2O = 1000 Pa; 1 mmHg = 1.36 cmH2O = 0.133 kPa = 133.3 Pa; 1 atm = 101.325 kPa = 760 mmHg = 1.013 bar; 1 bar = 100 kPa = 750 mmHg. [1]
- Distinguish absolute (referenced to vacuum), gauge (referenced to atmosphere, the usual clinical and cylinder reading) and differential pressure.
- Mechanical devices: Bourdon gauge (curved flattened tube, high-pressure cylinders), aneroid gauge (sealed capsule, anaesthetic-machine pressures), liquid manometer (mercury for BP, water for venous/airway, the calibration reference).
2. The transducer and the arterial line (4 marks)
- The invasive transducer converts hydraulic pressure into an electrical signal. Commonest design: a piezoresistive strain-gauge — fluid pressure deforms a thin diaphragm; four strain gauges bonded to it change resistance with deformation; arranged in a Wheatstone bridge they convert the change to a voltage proportional to pressure. (Mention the alternative capacitive design: diaphragm as one plate of a capacitor.)
- Arterial line components: arterial cannula (smaller diameters damp more; 20-gauge preferred for radial access); non-compliant pressure tubing; a continuous-flush device (pressurised bag of heparinised saline delivering a slow constant flush, the high driving head keeping flow independent of arterial pressure); the transducer; the monitor.
- Complications (briefly): temporary occlusion is commonest (1.5 to 35 per cent, mean about 20 per cent), major complications under 1 per cent and similar across radial, femoral and axillary sites; rarer serious entries are permanent ischaemic damage, sepsis, local infection, pseudoaneurysm, haematoma and bleeding; occlusion risk rises with catheter-to-vessel ratio, haematoma, duration of cannulation, and female sex [5].
References5ShowHide
- [1]Gardner RM Direct blood pressure measurement--dynamic response requirements. Anesthesiology, 1981.PMID 7469106
- [2]Kleinman B, Powell S, Kumar P The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology, 1992.PMID 1466471
- [3]Sessler DI Temperature monitoring and perioperative thermoregulation. Anesthesiology, 2008.PMID 18648241
- [4]Niven DJ, Gaudet JE, Laupland KB, et al Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis. Annals of Internal Medicine, 2015.PMID 26571241
- [5]Scheer B, Perel A, Pfeiffer UJ Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Critical Care, 2002.PMID 12133178