Anaes SAQsMeasurement & monitoring physics

Anaes SAQs · Measurement & monitoring physics

Pressure transduction, the arterial line, damping, and temperature measurement

A First-examination physics-and-measurement SAQ testing the full invasive-pressure and temperature-measurement framework: pressure as force per unit area with unit conversions; the strain-gauge (piezoresistive) transducer (diaphragm, strain gauges, Wheatstone bridge); the arterial line components (cannula, non-compliant tubing, pressurised heparinised-saline flush, transducer, monitor); zeroing to atmosphere and levelling to the mid-axillary line (right atrium) with the hydrostatic error of about 0.75 to 1 mmHg per cm; the natural frequency and damping (under/over/optimal) and the fast-flush square-wave test with what each state does to systolic, diastolic and mean; the four temperature-sensing principles (thermistor negative coefficient, thermocouple Seebeck voltage, RTD positive coefficient, infrared); and the monitoring sites (oesophageal/nasopharyngeal/pulmonary artery/bladder core vs skin/peripheral).

20 marks20 min4 min readVerification in progress

Target exams

ANZCAFRCAEDAICABAFCAI
On this page
Study tools

Target exams

ANZCAFRCAEDAICABAFCAI
Prompt
Outline the physical principles of invasive arterial pressure measurement. Describe the components of the arterial line, the requirements for zeroing and levelling, and the dynamic response of the system including how it is assessed. Briefly describe the four physical principles by which temperature is measured and the sites used for perioperative temperature monitoring. (20 marks)

Model answer plan (20 marks, 20 minutes)

A high-scoring answer is structured in five blocks that mirror the stem: (1) principles and units; (2) the transducer and the arterial line; (3) zeroing, levelling and the dynamic response; (4) temperature principles; (5) monitoring sites. Allocate roughly 4 marks to each block. [1]

1. Principles and units (4 marks)

  • Pressure is force per unit area (P = F/A); SI unit the pascal (Pa = N/m2).
  • Conversions (state at least three): 1 kPa = 7.5 mmHg = 10.2 cmH2O = 1000 Pa; 1 mmHg = 1.36 cmH2O = 0.133 kPa = 133.3 Pa; 1 atm = 101.325 kPa = 760 mmHg = 1.013 bar; 1 bar = 100 kPa = 750 mmHg. [1]
  • Distinguish absolute (referenced to vacuum), gauge (referenced to atmosphere, the usual clinical and cylinder reading) and differential pressure.
  • Mechanical devices: Bourdon gauge (curved flattened tube, high-pressure cylinders), aneroid gauge (sealed capsule, anaesthetic-machine pressures), liquid manometer (mercury for BP, water for venous/airway, the calibration reference).

2. The transducer and the arterial line (4 marks)

  • The invasive transducer converts hydraulic pressure into an electrical signal. Commonest design: a piezoresistive strain-gauge — fluid pressure deforms a thin diaphragm; four strain gauges bonded to it change resistance with deformation; arranged in a Wheatstone bridge they convert the change to a voltage proportional to pressure. (Mention the alternative capacitive design: diaphragm as one plate of a capacitor.)
  • Arterial line components: arterial cannula (smaller diameters damp more; 20-gauge preferred for radial access); non-compliant pressure tubing; a continuous-flush device (pressurised bag of heparinised saline delivering a slow constant flush, the high driving head keeping flow independent of arterial pressure); the transducer; the monitor.
  • Complications (briefly): temporary occlusion is commonest (1.5 to 35 per cent, mean about 20 per cent), major complications under 1 per cent and similar across radial, femoral and axillary sites; rarer serious entries are permanent ischaemic damage, sepsis, local infection, pseudoaneurysm, haematoma and bleeding; occlusion risk rises with catheter-to-vessel ratio, haematoma, duration of cannulation, and female sex [5].
Anaesthesia Pro

Continue reading

You have read the opening of this SAQ. The complete unit — every section and its primary-source references — is part of the Anaesthesia fellowship library.

Sign in to continueSee pricing
References5ShowHide
  1. [1]Gardner RM Direct blood pressure measurement--dynamic response requirements. Anesthesiology, 1981.PMID 7469106
  2. [2]Kleinman B, Powell S, Kumar P The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology, 1992.PMID 1466471
  3. [3]Sessler DI Temperature monitoring and perioperative thermoregulation. Anesthesiology, 2008.PMID 18648241
  4. [4]Niven DJ, Gaudet JE, Laupland KB, et al Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis. Annals of Internal Medicine, 2015.PMID 26571241
  5. [5]Scheer B, Perel A, Pfeiffer UJ Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Critical Care, 2002.PMID 12133178
Anaesthesia Pro

$29/ monthor $279 / year

The complete Anaesthesia library, every exam format, one subscription.

  • Complete library for one chosen specialty
  • Timed MCQ mock exams and spaced review
  • Clinical cases & cross-table vivas
See pricingSign in
PreviousPressure and temperature measurementMeasurement & monitoring physicsNextSI units and dimensional analysisMeasurement & monitoring physics