Anaes CasesThe anaesthetic machine

Anaes Cases · The anaesthetic machine

Oxygen supply failure during surgery

The classic anaesthetic-machine hot case — the pipeline oxygen failure (or the empty cylinder) during surgery. The discriminating management is: do NOT diagnose first; ventilate the patient by hand on 100 per cent oxygen from the reserve cylinder (open the cylinder valve, switch to the reservoir bag), call for help, switch to the self-inflating bag if the machine cannot deliver O2, and only then investigate the supply. The pre-use check that confirmed the reserve O2 cylinder was full and the fail-safe alarms worked is the single measure that prevents this becoming a hypoxic arrest.

hot5 min readVerification in progress

Target exams

ANZCAFRCAABAEDAICFCAIFCA_SA
On this page
Study tools

Target exams

ANZCAFRCAABAEDAICFCAIFCA_SA
Prompt
During a general anaesthetic under controlled ventilation, the oxygen pipeline pressure alarm fires, the inspired oxygen falls, the Bosun whistle sounds, and the ventilator bellows begins to collapse. The oxygen saturation is 94 per cent and falling. Manage.

Presentation

You are giving a general anaesthetic for an elective laparoscopic cholecystectomy in a healthy 45-year-old. The patient is paralysed and ventilated on a circle system with sevoflurane in oxygen and air. Suddenly the oxygen pipeline pressure alarm fires on the workstation, the inspired oxygen reading begins to fall, the Bosun whistle sounds from the machine, the ventilator bellows begins to collapse, and the oxygen saturation is 94 per cent and falling. The end-tidal CO2 trace is preserved. The pipeline pressure gauge on the machine reads zero. Manage.

Anaesthesia Pro

Continue reading

You have read the opening of this case. The complete unit — every section and its primary-source references — is part of the Anaesthesia fellowship library.

Sign in to continueSee pricing
References4ShowHide
  1. [1]Hartle A, Anderson E, Bythell V, et al. Checking anaesthetic equipment 2012. Anaesthesia, 2012.PMID 22563957
  2. [8]Goneppanavar U, Prabhu M. Anaesthesia machine: checklist, hazards, scavenging. Indian Journal of Anaesthesia, 2013.PMID 24249887
  3. [9]Mehta SP, Eisenkraft JB, Posner KL, Domino KB. Patient injuries from anesthesia gas delivery equipment: a closed claims update. Anesthesiology, 2013.PMID 23835591
  4. [11]Li J, Zhang Y, Gu W, et al. Diagnosis, management, and prevention of malfunctions in anesthesia machines. Technology and Health Care, 2023.PMID 37302057
Anaesthesia Pro

$29/ monthor $279 / year

The complete Anaesthesia library, every exam format, one subscription.

  • Complete library for one chosen specialty
  • Timed MCQ mock exams and spaced review
  • Clinical cases & cross-table vivas
See pricingSign in
PreviousA stuck unidirectional valve — CO2 rebreathingThe anaesthetic machineNextPipeline oxygen failure during surgeryThe anaesthetic machine