ICU Vivas · Mechanical ventilation
Ventilator waveforms: air-trapping and ineffective triggering
A CICM cross-table data viva on ventilator waveform interpretation. The candidate reads the flow and pressure scalars, recognises incomplete expiration and ineffective triggering, identifies intrinsic (auto-)PEEP on the expiratory hold, and explains the time-constant physiology and the management.
On this page
Study tools
Target exams
Stimulus
A 70-year-old man with severe COPD is intubated for a hypercapnic exacerbation. He is on volume-controlled ventilation. You are shown the ventilator scalar screen, an expiratory-hold manoeuvre, and a simultaneous arterial gas.[1]
The flow–time scalar: inspiratory flow is delivered as a constant upward deflection; the expiratory flow descends but does not return to the zero baseline before the next breath begins — the expiratory limb is truncated, with residual expiratory flow still present at the onset of the next inspiration.[1]
The pressure–time scalar: several small negative deflections appear below the baseline at irregular intervals that do not trigger a ventilator breath (ineffective efforts).[1]
The expiratory hold: a two-second end-expiratory occlusion records an airway pressure of 8 cmH2O above the set PEEP of 5 cmH2O (set PEEP 5; total PEEP 13).[1]
| Variable | Value |
|---|---|
| Mode | Volume control |
| Tidal volume | 420 mL |
| Set respiratory rate | 16 /min |
| FiO2 | 0.4 |
| Set PEEP | 5 cmH2O |
| Measured intrinsic PEEP | 8 cmH2O |
| Peak inspiratory pressure | 28 cmH2O |
| pH | 7.32 |
| PaCO2 | 58 mmHg |
| PaO2 | 78 mmHg |
| Bicarbonate | 30 mmol/L |
Take ninety seconds, then present a structured interpretation.[1]
References2ShowHide
- [1]Bersten AD, Soni N (eds). Oh's Intensive Care Manual. 8th edition, Elsevier, 2018.
- [2]Marino PL. The ICU Book. 4th edition, Lippincott Williams & Wilkins, 2014.