llms.txtPsychiatry LLM catalogSitemap

Anaes CasesElectricity, diathermy & theatre safety

Anaes Cases · Electricity, diathermy & theatre safety

The poorly applied return plate and the burn risk

The short case — the return-plate burn from the poor contact (the bony prominence, the small area). The current density at the plate rises when the contact is poor; the alternate-site burns (the ECG electrode) occur when the plate is detached and the current returns through another path.

short2 min readVerification in progress

Target exams

ANZCAFRCAABAEDAIC
On this page
Study tools

Target exams

ANZCAFRCAABAEDAIC
Prompt
After a long case, a burn is found under the diathermy return plate site. The plate had been applied over a bony prominence and the contact-quality monitor had alarmed intermittently. Explain the mechanism and the prevention.

Presentation

After a three-hour laparotomy, a 4 by 4 cm partial-thickness burn is found on the patient's hip, beneath the diathermy return plate. The plate had been applied over the greater trochanter (a bony prominence), and the contact-quality monitor had alarmed intermittently during the case. The surgeon used the monopolar diathermy extensively. Explain the mechanism and the prevention.[34181253]

Anaesthesia Pro

Continue reading

You have read the opening of this case. The complete unit — every section and its primary-source references — is part of the Anaesthesia fellowship library.

Sign in to continueSee pricing
References2ShowHide
  1. [2]Link T. Guidelines in practice: electrosurgical safety. AORN Journal, 2021.PMID 34181252
  2. [34181253]Bersten AD, Soni N (eds) Oh's Intensive Care Manual. 8th edition, Elsevier, 2018.
Anaesthesia Pro

$29/ monthor $279 / year

The complete Anaesthesia library, every exam format, one subscription.

  • Complete library for one chosen specialty
  • Timed MCQ mock exams and spaced review
  • Clinical cases & cross-table vivas
See pricingSign in